IEEE TRANSACTIONS AND JOURNALS TEMPLATE 1
MonoLoT: Self-Supervised Monocular Depth
Estimation in Low-Texture Scenes for Automatic
Robotic Endoscopy
Qi He, Guang Feng, Sophia Bano, Danail Stoyanov, Siyang Zuo
Abstract— The self-supervised monocular depth estima-
tion framework is well-suited for medical images that lack
ground-truth depth, such as those from digestive endo-
scopes, facilitating navigation and 3D reconstruction in the
gastrointestinal tract. However, this framework faces sev-
eral limitations, including poor performance in low-texture
environments, limited generalisation to real-world datasets,
and unclear applicability in downstream tasks like visual
servoing. To tackle these challenges, we propose MonoLoT,
a self-supervised monocular depth estimation framework
featuring two key innovations: point matching loss and
batch image shuffle. Extensive ablation studies on two
publicly available datasets, namely C3VD and SimCol, have
shown that methods enabled by MonoLoT achieve sub-
stantial improvements, with accuracies of 0.944 on C3VD
and 0.959 on SimCol, surpassing both depth-supervised
and self-supervised baselines on C3VD. Qualitative eval-
uations on real-world endo

## B0002 · 第1页 · 清洗文本字符1201–2400

scopic data underscore the
generalisation capabilities of our methods, outperforming
both depth-supervised and self-supervised baselines. To
demonstrate the feasibility of using monocular depth es-
timation for visual servoing, we have successfully inte-
grated our method into a proof-of-concept robotic platform,
enabling real-time automatic intervention and control in
digestive endoscopy. In summary, our method represents
a significant advancement in monocular depth estimation
for digestive endoscopy, overcoming key challenges and
opening promising avenues for medical applications.
Index Terms — Monocular depth estimation, automatic
intervention, digestive endoscopy.
I. I NTRODUCTION
E
NDOSCOPY has long served as the golden standard
investigative procedure for clinical diagnosis, providing
high-definition internal tissue images with a narrow scope.
During the diagnostic process, clinicians record anatomical
landmarks following specific guidelines, providing sources for
This work was supported by the National Natural Science Foundation
of China under Grant No. 62133010, Wellcome/EPSRC Centre for
Interventional and Surgical Sciences (WEISS), UK (203145/Z/16/Z),
and the Engineering a

## B0003 · 第1页 · 清洗文本字符2401–3600

nd Physical Sciences Research Council (EPSRC)
(EP/P027938/1). (Qi He and Guang Feng contribute equally to the
article. Corresponding author: Siyang Zuo.)
Qi He, Guang Feng, and Siyang Zuo are with the Key Laboratory
of Mechanism Theory and Equipment Design of Ministry of Education,
Tianjin University, Tianjin 300354, China (e-mail: howard@tju.edu.cn;
fengguang@tju.edu.cn; siyang zuo@tju.edu.cn).
Sophia Bano, Danail Stoyanov are with the Wellcome/EPSRC Centre
for Interventional and Surgical Sciences (WEISS) and the Department
of Computer Science, University College London, London W1W 7TY ,
UK. (e-mail: sophia.bano@ucl.ac.uk; danail.stoyanov@ucl.ac.uk).
developing an image recognition system [1]. On the other
hand, computer graphics techniques, including 3D reconstruc-
tion and monocular depth estimation, provide an alternative
approach to locating these landmarks [2], [3]. Furthermore,
robot platforms based on traditional endoscopes or capsule
endoscopes have shown more comfortable and less invasive
alternative approaches [4], [5].
Depth estimation is crucial for downstream clinical diag-
nosis tasks, such as navigation [6], [7] and active stabilisa-
tion [8], [9], and therefore, it

## B0004 · 第1页 · 清洗文本字符3601–4800

 is the primary focus of this
study. The monocular depth estimation method calculates the
spatial distance between the viewpoint and corresponding 3D
points using a single camera and associated algorithms. The
monocular camera model is applicable to digestive endoscopy.
However, previous methods face several challenges, including
underperformance in low-texture scenes, limited adaptability
to real-world datasets, and unclear applicability of depth
estimation in visual servoing tasks.
Notably, this task lacks ground-truth (GT) depth in real-
world scenarios due to the absence of depth sensors, such as
depth cameras, in digestive endoscopy, making self-supervised
training prevalent in this field. However, the low texture
present in endoscopic images poses a significant challenge
for self-supervised monocular depth estimation methods [10]–
[12], as such scene result in inaccurate photometric recon-
struction errors, thereby impeding the training process. Re-
cent advances have explored the utilisation of Generative
Adversarial Networks (GANs) to acquire depth benchmarks
from alternative image domains like chromoendoscopy [2],
and simulators [13]. But these methods demonstrate limited


## B0005 · 第1页 · 清洗文本字符4801–6000

generalisability to real endoscopy images and rely on external
chromoendoscopy images or synthetic data during training.
Moreover, achieving real-time performance with minimal la-
tency is crucial for visual servoing applications, making a
lightweight model during inference more desirable than a cost
volume-based model [14].
Recent efforts to address low-texture challenges, including
the Transformer-based local feature matching method LoFTR
[15], have shown potential in generating high-quality matches.
We integrated this feature-matching method into our training
process to enhance the self-supervised monocular depth esti-
mation frameworks for low-texture scenes. Furthermore, the
availability of publicly accessible endoscopic datasets, such
as C3VD [16], SimCol [17], EndoSLAM [3], Endomapper
[18], and upper gastrointestinal (GI) dataset [19], has expanded
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistrib

## B0006 · 第1页 · 清洗文本字符6001–6253

ution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0007 · 第2页 · 清洗文本字符1–1200

2 IEEE TRANSACTIONS AND JOURNALS TEMPLATE
PharynxSquamocolumnar JunctionMiddle-upper bodyLower bodyAntrum
User interface Global coordinate system view Camera view
Fig. 1. Partial experimental frames of automatic robotic endoscopy on
a realistic upper GI tract phantom. The camera poses and trajectories
obtained from the Aurora tracker are depicted in red (current frame) and
blue (previous frames) pyramid. Full details is provided in Video S1.
significantly. These datasets, encompassing phantom data,
synthetic data, and real-world data, serve as valuable resources
for training and validating models.
In this paper, we present a novel approach to address
the challenges posed by inaccurate photometric loss during
the training of self-supervised monocular depth estimation
methods in low-texture scenes. We propose the point-matching
loss, refining reprojected points of the monocular depth model
through the feature correspondences derived from a feature-
matching model (such as LoFTR). The efficiency of this step
is improved by the adaptive weight and offline preprocessing.
Additionally, the batch image shuffle loss is proposed to
discriminate similar patches with different depths. Extensi

## B0008 · 第2页 · 清洗文本字符1201–2400

ve
experiments demonstrate the effectiveness of our proposed
methods in overcoming these challenges. Furthermore, we
present a proof-of-concept robotic platform to illustrate the
feasibility of using monocular depth estimation models for
automatic robotic endoscopy (see Fig. 1).
Our paper contributes significantly to the field of self-
supervised monocular depth estimation in GI endoscope
videos, with the key highlights as follows:
(i) We introduce a novel objective function that incorpo-
rates the point matching loss and batch image shuffle
loss, effectively enhancing the performance of the self-
supervised Monocular depth estimation in Low-Texture
scenes (MonoLoT). Our methods resulted in improved
accuracy (θ <1.25) from 4.7% to 18.8% on the C3VD
dataset and from 4.4% to 5.7% on the SimCol dataset.
(ii) We present comprehensive quantitative and qualitative
results on the C3VD dataset, SimCol dataset, upper-GI
dataset, EndoSLAM dataset, and EndoMapper dataset,
demonstrating the effectiveness and robustness of our
proposed methods.
(iii) We have successfully integrated monocular depth estima-
tion into a robotic platform designed for visual servoing.
Our approach is thoroughly vali

## B0009 · 第2页 · 清洗文本字符2401–3600

dated on a realistic GI
phantom, confirming its robustness and practical appli-
cability in real-world scenarios.
By combining these contributions, our work significantly
advances the field of self-supervised monocular depth esti-
mation, offering valuable insights for both researchers and
practitioners in the field of visual servoing.
II. R ELATED WORK
A. Depth Estimation for Monocular Endoscope
Depth estimation of monocular endoscopic images has been
an important research topic in recent years. Various approaches
have been proposed to tackle this task. For example, Rau
et al. [13] introduced a method based on GANs to estimate
depth maps from monocular endoscopic images. Their method
transfers the network’s monocular depth estimation capabilities
from the synthetic image domain to real endoscopic images.
In another notable work by Widya et al. [20], the focus was on
addressing the issue of balancing weights between the depth
and pose loss terms by employing unsupervised photometric
error as a general term of the supervised depth and pose error.
Rodriguez-Puigvert et al. [21] presented a Bayesian deep net-
work approach to explicitly model data uncertainty and model
uncertainty for

## B0010 · 第2页 · 清洗文本字符3601–4800

 single-view depth estimation in colonoscopies.
Their study involved ensembling multiple deep learning-based
depth estimation models as a Bayesian teacher model, enabling
the student model to learn from a similar distribution. In
summary, the existing literature lacks comprehensive research
on addressing the challenges posed by low-texture regions in
endoscopic images, which is the primary focus of this paper.
B. Feature Extraction in Low-Texture Regions
Low-texture regions pose a common challenge in both
indoor scenes and GI scenarios. To address this issue, several
studies have investigated specific prior or global attention
schemes. For instance, Li et al. [22] introduced a co-planar
constraint specifically designed for low-texture regions in
indoor images. Their method utilised geometric and colour
dissimilarities to constrain the estimated depth, enabling re-
liable depth estimation in challenging regions lacking rich
textures. Like Saunders et al. [23], our method also bypass the
homography warping to input augmented images. However,
our starting point is different because their baseline model
was able to achieve highly accurate depth maps on augmented
images. Therefore, they

## B0011 · 第2页 · 清洗文本字符4801–5596

 utilised the correspondence between
unaugmented and augmented images, using the unaugmented
depth to supervise the augmented depth, thereby improving
the accuracy on augmented images. In contrast, our baseline
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0012 · 第3页 · 清洗文本字符1–1200

QI et al.: MONOLOT: SELF-SUPERVISED MONOCULAR DEPTH ESTIMATION IN LOW-TEXTURE SCENES FOR AUTOMATIC ROBOTIC ENDOSCOPY 3
(𝐼𝑎, 𝐼𝑏)
Depth and pose
branch
Photometric
reconstruction branch
Point matching
branch
Batch image shuffle
branch
(𝐼𝑎, 𝐼𝑏) (𝐼𝑎, 𝐼𝑏) (𝐼𝑎, 𝐼𝑏) 𝐼𝑎
𝑑𝑎
𝑅 | 𝑇 𝑎→𝑏
𝑝𝑎→𝑏
𝑑𝑎
ℒ𝑠
𝑑𝑎
𝜃𝑝
𝜃𝑑
∇ 𝑜𝑎 VI
𝑜𝑏
𝑅 | 𝑇 𝑎→𝑏
∇
𝐼𝑏
𝐼𝑎
𝐼𝑎
ℒ𝑚
(𝐼𝑎, 𝐼𝑏)
𝜃𝑚
𝑝𝑎→𝑏 Ƹ𝑝𝑎
Indexing
𝑝
𝑥
𝑧
𝑦 𝑥′
𝑦′
𝑧′
View b
𝑜𝑎
𝑜𝑏
View a
(c) Homo-Warp
(f) Fixed batch image shuffle branch
Horizontal batch image shuffle Horizontal batch image shuffle
Vertical batch image shuffle Vertical batch image shuffle
𝑑𝑏𝑖𝑠 𝑑𝑟𝑒𝐼𝑏𝑖𝑠
𝑝𝑑𝑖𝑟
1 − 𝑝𝑑𝑖𝑟
𝑑𝑏𝑖𝑠 𝑑𝑟𝑒𝐼𝑏𝑖𝑠
ℒ𝑏𝑖𝑠
†𝑑𝑎
∇∇∇
∇ ∇ ∇
𝜃𝑑
𝜃𝑑
( Ƹ𝑝𝑎, Ƹ𝑝𝑏)
(d) Depth and pose branch (e) Point matching branch
ℒ𝑟
𝐼𝑏→𝑎
𝑝𝑎→𝑏 Sampling
𝑝𝑎
∇ ∇ ∇
Homo-Warp
𝐷𝑎 𝑅 | 𝑇 𝑎→𝑏
𝐼𝑏 𝐼𝑎
(b) Photometric reconstruction branch
𝑑𝑏𝑖𝑠 𝑑𝑟𝑒𝐼𝑏𝑖𝑠
(g) Random batch image shuffle branch
𝐼𝑎
Random batch image shuffle Random batch image shuffleShared random areas
∇ ∇ ∇ ℒ𝑏𝑖𝑠
∗
𝑑𝑎
𝜃𝑑
∇
𝐼𝑎
∇
(a) MonoLoT architecture
Fig. 2. Monocular depth estimation pipeline. The overall architecture is depicted in (a). The fundamental monocular depth estimation pipeline
comprises (b) the photometric reconstruction branch, (c) homography warping, and (d)

## B0013 · 第3页 · 清洗文本字符1201–2400

 the depth and pose branch, highlighted within blue dash
rectangles. The proposed methods, such as (e) the point matching branch, (f) the fixed batch image shuffle branch, and (g) the random batch image
shuffle branch, are emphasised within red dash rectangles. Gradient back-propagation is depicted by the red arrows.
model performs poorly on unaugmented images. Zhao et al.
[24] proposes the use of COLMAP to estimate preliminary
poses, followed by neural network optimisation of the pose,
specifically targeting the characteristic of large rotational
movements between adjacent frames in indoor videos. How-
ever, GI videos do not exhibit such large rotations as seen
in indoor scenes. Moreover, due to their weak texture char-
acteristics, the performance of pose estimation by COLMAP
tends to be poor for GI videos [17]. Unlike [25], which
utilises the depth from the previous epoch model for self-
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is pe

## B0014 · 第3页 · 清洗文本字符2401–2689

rmitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0015 · 第4页 · 清洗文本字符1–1200

4 IEEE TRANSACTIONS AND JOURNALS TEMPLATE
distillation of the current model, our method constructs a
new regularisation term and does not block the gradient from
back-propagating along the pseudo-label depth. Furthermore,
Sun et al. [15] proposed a Transformer-based approach for
generating dense matches in low-texture areas. By extracting
coarse-to-fine interest points in global receptive field, their
method improved the feature-matching performance in low-
texture regions. Since the digestive tract is mostly composed
of curved surfaces, the coplanar constraint is not suitable for
the digestive tract scene in our study. Therefore, we instead
focus on extracting feature points in weak textures to improve
self-supervised training.
C. Transformer in Depth Estimation
In recent studies, Transformer-based models [26] have
been explored for monocular endoscopy depth estimation,
showcasing their potential in this domain [11], [12]. For
instance, MonoViT [11] replaced the convolutional neural
network (CNN)-based depth encoder of Monodepth2 [10] with
MPViT [27], while Lite-Mono [12] employed consecutive
dilated convolutions [28] and cross-covariance attention [29]
to strengthen their depth e

## B0016 · 第4页 · 清洗文本字符1201–2400

ncoder. In our study, we validate
the limitations of these models on digestive endoscopy image
data and construct suitable general modules to improve their
performance.
III. M ONOCULAR DEPTH ESTIMATION IN LOW-TEXTURE
ENDOSCOPIC VIDEOS
We observed that the poor performance of self-supervised
monocular depth estimation baseline is primarily due to in-
accurate photometric reconstruction errors. To address this
issue, the point matching loss ( Lm) and batch image shuffle
loss ( L†
bis for fixed and L∗
bis for random) are proposed and
introduced in the following sections.
A. Framework Overview
We present a summary of the state-of-the-art self-supervised
monocular depth estimation pipeline [10]–[12], as seen in Fig.
2(b)(c)(d). Initially, PoseNet θp and DepthNet θd predict the
viewpoint [R | T ]a→b and inverse depth da from the image
pair (Ia, Ib) using the depth and pose branches (see Fig. 2(d)).
Subsequently, the synthesised frame Ib→a is generated by
the photometric reconstruction branch (see Fig. 2(b)). Their
objective function consists of two key components:
L = µLr + λLs, (1)
(i) Lr (Minimum Reprojection Loss) : this term aims to
minimise the photometric reconstruction error pe(·)

## B0017 · 第4页 · 清洗文本字符2401–3600

 and
is defined as:
Lr = min
b
pe(Ia, Ib→a). (2)
(ii) Ls (Edge-Aware Smoothness Loss) : this term is de-
signed to smooth the depth maps and is expressed as:
Ls = |∂xd∗
a|e−|∂xIa| + |∂yd∗
a|e−|∂yIa|. (3)
Here, d∗
a = da/da denotes the mean-normalised inverse
depth [10].
Additionally, µ is used for ”Auto-Masking” [10], calculated
as:
µ = min
b
pe(Ia, Ib→a) < min
b
pe(Ia, Ib), (4)
where λ plays a role in determining the relative importance of
Lr and Ls in the overall loss.
In our proposed objective function, which extends the state-
of-the-art pipeline in (1), we introduce two new components:
L = µLr + λLs + Lm + Lbis. (5)
(i) Lm (Point Matching Loss) : this loss is formulated to
refine the reprojected points by incorporating pseudo
label supervision during training.
(ii) Lbis (Batch Image Shuffle Loss): this loss is designed to
enhance the model’s ability to generalise to unseen data
in low-texture scenes by training on synthetic images
with similar textures.
The following paragraphs will provide a detailed expla-
nation of the Point Matching Loss and the Batch Image
Shuffle Loss, starting from the fundamentals of the minimum
reprojection loss, also known as photometric reconstructi

## B0018 · 第4页 · 清洗文本字符3601–4800

on
error.
B. Photometric Reconstruction Error
Given current image frame Ia, the image frame Ib of view-
point [R | T ]a→b, current monocular depth Da, and the camera
intrinsic matrix K, the photometric reconstruction error loss
pe(·) gathers the differences between Ia and synthesised frame
Ib→a from [R | T ]a→b:
pe(Ia, Ib→a) = (1−α)∥Ia, Ib→a∥1 + α
2 (1−SSIM(Ia, Ib→a)),
(6)
where SSIM(·) is the structure similarity loss [30] and α =
0.85 following the setting from [10]. Here Ib→a denotes the
value obtained by sampling Ib using reprojected points pa→b
on the pixel coordinates system of Ib:
Ib→a = Ib⟨pa→b⟩, (7)
and pa→b = K · (Ra→b · (K −1 · pa · Da) + Ta→b). (8)
Here ⟨·⟩ denotes a sampling operator and pa signifies the target
location in the pixel coordinate system of Ia.
To summarise, the loss termLr is back-propagated to update
the parameters of Ra→b, Ta→b, and Da (as depicted in Fig.
2(b)). These parameters are obtained through PoseNet θp and
DepthNet θd (as illustrated in Fig. 2(d)), calculated as:
[R | T ]a→b = θp(Ia, Ib), (9)
Da = 1
( 1
maxD
+ ( 1
minD
− 1
maxD
)da)
, (10)
where, inverse depth da in the range of [0, 1] is transformed
into depth within the range of [minD, maxD].

## B0019 · 第4页 · 清洗文本字符4801–5543

 In this case,
the parameters θp and θd are primarily obtained through self-
supervised learning by minimising the photometric reconstruc-
tion error in (6).
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0020 · 第5页 · 清洗文本字符1–1200

QI et al.: MONOLOT: SELF-SUPERVISED MONOCULAR DEPTH ESTIMATION IN LOW-TEXTURE SCENES FOR AUTOMATIC ROBOTIC ENDOSCOPY 5
However, there is a special case in which when the location
of the reprojected point pa→b lies in a low-texture area, the
sampling operation in (7) will sample from pixels with very
similar colours. This ultimately results in multiple minima for
the term ∥Ia, Ib→a∥1 of the photometric reconstruction errors
in (6), thereby weakening its supervisory effect.
C. Point Matching Loss
In response to this challenge, we have introduced the Point
Matching Loss (PM) to improve the accuracy of reprojected
points in low-texture regions.
1) Core mechanism: In addition to minimising the photo-
metric reconstruction error, we can optimise θd and θp by
reducing the distance between the reprojected points pa→b
and their corresponding feature points ˆpb on Ib, as determined
by the local feature matching model (as shown in Fig. 2(e)).
To elaborate, when given the image pair (Ia, Ib), the local
feature matching model θm identifies corresponding feature
points (ˆpa, ˆpb) as follows:
(ˆpa, ˆpb) = θm(Ia, Ib), (11)
where θm was pretrained and remained frozen during our
training process. In

## B0021 · 第5页 · 清洗文本字符1201–2400

 practice, we chose LoFTR [15] as our
θm due to its outstanding feature point extraction ability
on challenging scenarios such as texture-less wall and floor
with repetitive patterns, which are commonly encountered in
endoscopic videos. The point-matching process is illustrated
in Fig. 3, highlighting the matching ability of LoFTR.
Subsequently, we substituted pa in (8) with ˆpa and min-
imised the Point Matching Loss Lm between the reprojected
feature point ˆpa→b from Ia to Ib and the corresponding feature
point ˆpb on Ib:
Lm = ∥ˆpb, ˆpa→b∥1. (12)
Here, ˆpa→b is computed as:
ˆpa→b = K · (Ra→b · (K −1 · ˆpa · Da) + Ta→b). (13)
It’s worth noting that ˆpb consists of subpixel (or floating-point
coordinates), while ˆpa uses integer coordinates, which can also
serve as indices into tensors, denoted as [ˆpa]. Consequently,
instead of passing ˆpa to the homography warping module, we
directly employed ˆpa as the index for pa and pa→b (see Fig.
2(e)). Therefore, equation (13) can be simplified as:
ˆpa→b = pa→b[ˆpa], (14)
and pa→b[ˆpa] = K · (Ra→b · (K −1 · pa[ˆpa] · Da) + Ta→b).
(15)
2) Adaptive Point Matching Loss: While the point-matching
loss is expected to decrease with the optimisatio

## B0022 · 第5页 · 清洗文本字符2401–3600

n of the
objective function, the inference error of the local feature-
matching model can impede this process, especially when the
photometric reconstruction error decreases below the point-
matching loss. Specifically, the corresponding feature points
ˆpb acquired through the local feature matching model θm may
not be optimal, and we posit that the error between these points
𝐼𝑎 𝐼𝑏 Ƹ𝑝𝑎 Ƹ𝑝𝑏
Fig. 3. Visualisation of LoFTR-based Point Matching. Compared with
the general scene, the corresponding feature points ( ˆpa and ˆpb) are
harder to detect (highlighted with blue dots) and pair (highlighted with
green lines) in GI endoscopic images (Ia and Ib) due to the low-texture
feature. Fortunately, this method can handle these task robustly.
and the reprojected points ˆpa→b remains within the predefined
threshold δ. We define the Adaptive Point Matching Loss as:
Lm =
(
∥ˆpb, ˆpa→b∥1 · eps if ∥ˆpb, ˆpa→b∥1 < δ
∥ˆpb, ˆpa→b∥1 otherwise (16)
where eps is set to 10−4.
3) Improving Computational Efficiency via Offline Prepro-
cessing: In practice, we found that our point-matching loss
module significantly increased training time, tripling it, leading
to an approximately 9 GB increase in GPU memory

## B0023 · 第5页 · 清洗文本字符3601–4800

 usage.
This was primarily due to the computationally intensive and
GPU memory-demanding inference of (11). To address this
issue, we now calculate the corresponding feature points for
each image pair as a preprocessing step before training. In this
revised approach, the point matching loss not only no longer
extends the training time but also requires a minimal additional
amount of GPU memory to execute.
D. Batch Image Shuffle Loss
Another challenge in self-supervised training is data aug-
mentation, especially in low-texture endoscopic scenes, where
pixel blocks may have uniform textures but represent varying
depths. Specifically, if image frames Ia and Ib are related by
the homography warping, considering the homography warp-
ing of augmented content is necessary when using augmented
images to compute the photometric reconstruction error loss
(Fig. 2(b)). However, bypassing the photometric reconstruc-
tion error allows us to perform image augmentation without
considering the homography warping. We propose the Batch
Image Shuffle Loss , which optimises the objective function
through self-supervised learning by directly comparing the
inverse depths before and after image augmentat

## B0024 · 第5页 · 清洗文本字符4801–6000

ion. We regard
the characteristic of low texture as having similar textures
but different depths, and we need the model make correct
predictions for such image areas. Through batch image shuffle,
we can construct such regions at the junction of stitched
images, and encourage the model to make correct predictions
through self-supervision. As shown in Fig. 4, MonoLoT makes
more accurate predictions in these kinds of junction areas
compared to the baseline.
Input batch, denoted as Ia ∈ RF ×3×H×W , is divided into
upper and lower parts along the first dimension ( F ). Swap the
top halves of the two half-batches along the third dimension
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE X

## B0025 · 第5页 · 清洗文本字符6001–6026

plore. Restrictions apply.

## B0026 · 第6页 · 清洗文本字符1–1200

6 IEEE TRANSACTIONS AND JOURNALS TEMPLATE
Input
GT depth
Baseline 𝐷𝑟𝑒
MonoLoT 𝐷𝑟𝑒
Shuffled input
GT depth
Baseline 𝐷𝑏𝑖𝑠
MonoLoT 𝐷𝑏𝑖𝑠
Fig. 4. Visualisation of Batch Image Shuffle. The first column illustrates the input (where batch size is 2 for exemplification). The second column
shows the corresponding GT depths. Taking the horizontal batch image shuffle as an example, the third column shows the shuffled images ( Ibis)
after the first shuffle, followed by the corresponding GT depths in the fourth column. The fifth and sixth columns show the shuffled depth generated
by baseline and MonoLoT respectively. Predicted depths in the seventh and eighth columns shows MonoLoT outperforms the baseline by producing
sharper, more detailed depth maps in junction areas (highlighted with yellow boxes).
(H) and merge them to form the shuffled input batch:
shufflever(Ia) =
Ia[ F
2 :, :, : H
2 , :] Ia[: F
2 , :, H
2 :, :]
Ia[: F
2 , :, : H
2 , :] Ia[ F
2 :, :, H
2 :, :]

. (17)
Likewise, horizontal batch image shuffling, as represented by:
shufflehor(Ia) =

Ia[ F
2 :, :, :, : W
2 ] Ia[: F
2 , :, :, W
2 :]
Ia[: F
2 , :, :, : W
2 ] Ia[ F
2 :, :, :, W
2 :]

, (18)
allows for batch image shuffling a

## B0027 · 第6页 · 清洗文本字符1201–2400

long the fourth dimension
(W ). Introduce a probability term, allowing the model to use
vertical batch image shuffling with a probability of pdir (see
Fig. 2(f)). Once the shuffled input batch is constructed, it is fed
to the DepthNet, which estimates the shuffled inverse depth.
This shuffled inverse depth is then rearranged to generate the
reordered inverse depth dre, denoted as:
dre =
(
shufflever(θd(shufflever(I))) w.p. pdir
shufflehor(θd(shufflehor(I))) w.p. 1 − pdir
(19)
Minimise the batch image shuffle loss between the inverse
depth of the original batch and the reordered batch:
Lbis = ∥d, dre∥1, (20)
In addition to this fixed batch image shuffling approach, we
have explored further randomisation of the swapped areas. As
depicted in Fig. 2(g), random patches of the two half-batches
are swapped before and after depth estimation. The randomly
generated areas share the same location and size, ensuring
consistent illumination. It’s worth noting that we ensure the
randomly generated coordinates are divisible by 8 at scale 0
to ensure precise alignment with the coordinates on scales 1,
2, and 3. We formally denote the Fixed Batch Image Shuffle
loss (FBIS) as L†
bis, and the Random 

## B0028 · 第6页 · 清洗文本字符2401–3600

Batch Image Shuffle loss
(RBIS) as L∗
bis.
IV. R OBOTIC GASTROINTESTINAL INTERVENTION :
STRUCTURE AND DEPTH -BASED CONTROL
To demonstrate the robustness of the proposed MonoLoT
on a realistic robotic endoscopy setup, we have implemented
real-time visual servoing on a real GI endoscopy system
using our proof-of-concept robotic platform. This achievement
demonstrates the applicability of our proposed monocular
depth estimation method and the potential for utilising depth-
based control in automatic intervention.
A. Robotic platform
The proposed automatic intervention method is imple-
mented and validated using a novel robotic platform currently
under development at Tianjin University laboratory. For com-
pleteness, we have included necessary details for the robotic
platform sufficient to demonstrate the robustness of MonoLoT
in practical application (see Fig. 5). It consists of two primary
components:
6-DOF Serial Manipulator : This manipulator is responsi-
ble for adjusting the initial pose of the endoscopy. It provides
six joints for precise control. The sixth joint (motor 4) of the
manipulator allows rotation along the z-axis of the camera
coordinates system.
End-effector: The end

## B0029 · 第6页 · 清洗文本字符3601–4800

-effector is affixed to the sixth joint
of the manipulator. Motor 1 and 2 control the rotation of the
small knobs and large knobs separately, enabling the bending
of the endoscope along the x-axis and y-axis of the camera
coordinate system. Motor 3 is responsible for the insertion and
retraction of the endoscope.
To simulate gastroscopy, we integrated the robotic platform
with a GI endoscopy system (CV-290, Olympus, Japan) and
utilised an upper GI tract phantom made of soft rubber material
(see Fig. 5(b)). This phantom closely simulates the actual
condition within the human GI tract, providing a realistic
environment for testing and experimentation. To ensure precise
pose collection for evaluation purposes, a real-time electro-
magnetic sensor (Aurora, NDI, Canada) was incorporated into
the working channel of the endoscope.
B. Depth-Based Control
The control framework for automatic intervention consists
of three parts (see Fig. 5(c)). The endoscope captures videos of
the upper GI tract, transmitting them to image processing mod-
ules via a video capture card. Captured images are processed to
convert them into depth maps using MonoLoT-enabled model.
Within the depth map, the deepest

## B0030 · 第6页 · 清洗文本字符4801–5411

 point is selected as the
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0031 · 第7页 · 清洗文本字符1–1200

QI et al.: MONOLOT: SELF-SUPERVISED MONOCULAR DEPTH ESTIMATION IN LOW-TEXTURE SCENES FOR AUTOMATIC ROBOTIC ENDOSCOPY 7
Depth estimation
𝜃𝑑
Video capture
Compute anchor
PID
Controller
(c) Perception and control
Endoscope
Upper GI robotic platform
𝐸𝑥𝑦
𝐼𝑎
𝐷𝑎
𝐼𝑎
𝑂𝑥𝑦
𝐸𝑥𝑦
User interface
(a) Robotic platform x
y
z
1
2
3
4
5
2 3
4 5
12
3 5
4
Motor 1
Motor 2
Motor 3
Motor 4
Camera
1
2
3
4
5
1
Aurora
Tracker
Upper GI tract model
(b) Phantom
x
y
z
Global coordinate system
𝑣𝑥𝑦
𝑣𝑧
𝐼𝑎
𝐷𝑎
𝐸𝑥𝑦
𝐴𝑥𝑦
Compute
velocity
Fig. 5. Overview of Robotic Platform. (a) The robotic platform comprises four motors that manipulate the camera’s pose. (b) The phantom is
equipped with the Aurora sensor for pose tracking and recording. (c) Perception and control integrate our depth estimation into a PID controller,
enabling real-time visual servoing. The global coordinate system is defined to maintain consistency with the Aurora system: +x points inward, +y
points to the left, +z points upward.
TABLE I
C3VD S PLIT. THE INDEX OF EACH VIDEO IS FORMATTED AS "<T EXTURE ><V IDEO >" [16].
Cecum Descending Sigmoid Transcending
Index 1a 1b 2a 2b 2c 3a 4a 4b 4a 1a 2a 3a 3b 1a 1b 2a 2b 2c 3a 3b 4a 4b
Train 276 765 370 1,142 595 

## B0032 · 第7页 · 清洗文本字符1201–2400

730 700 514 61 700 194 103 235 250 214
Val 465 613 382
Test 425 148 536 597
TABLE II
SIMCOL SPLIT. THE INDEX OF EACH VIDEO IS DERIVED FROM THE SIMCOL CHALLENGE [17].
SyntheticColonI SyntheticColon II SyntheticColon III
Index S1-S3 S4 S5 S6-S8 S9 S10 S11-S13 S14 S15 B1-B3 B4 B5 B6-B8 B9 B10 B11-B13 B14 B15 O1-O3
Train 3,603 3,603 3,603 3,603 3,603 3,603
Val 1,201 1,201 1,201 1,201 1,201 1,201
Test 1,201 1,201 1,201 1,201 1,201 1,201 1803
anchor point, denoted as Axy = arg maxxy(Da). The target
direction − →O is calculated as:
− →O = Axy − Cxy
Cxy
, (21)
where Cxy =
 W/2 H/2
represents the image centre. The
anchor point coordinates, combined with the current state of
motor 1 and 2, serve as target angles Oxy for the robotic
platform’s controller, calculated as:
Oxy = Exy + s · − →O , (22)
where, Exy =
 Motor1 Motor2

denotes the current angles
of motor 1 and 2 from the motor encoder, while s signifies
the scale of each move. Utilising the received Oxy from the
image processing modules, the robotic platform adjusts the
rotation of the large knob and the small knob to minimise − →O .
This adjustment is performed using the proportional-integral-
This article has been accepted for publ

## B0033 · 第7页 · 清洗文本字符2401–2946

ication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0034 · 第8页 · 清洗文本字符1–1200

8 IEEE TRANSACTIONS AND JOURNALS TEMPLATE
derivative (PID) controller:
vxy = PID(Oxy, Exy, dt), (23)
where vxy is the manipulated velocity of motor 1 and 2, dt
is the time interval of our PID control. It’s worth noting that
our PID controller operates at a different frequency from the
image processing modules, necessitating separate readings of
the motor state. The speed control of insertion and retraction
is determined as:
vz = maxvz
minvz
− ∥− →O ∥2√
2
· maxvz . (24)
Here, vz denotes the velocity of motor 3, which falls within
the predefined range of [minvz , maxvz ]. In implementation,
position limits are set for motors 1, 2, and 3 to ensure safety
throughout the entire intervention. Additionally, a moving
average is applied on the anchor point to ensure a smooth
transition between targets.
In summary, this control framework relies on real-time
monocular depth estimation, enabling the robotic platform
to accurately adjust the endoscope’s position and reach the
desired target within the upper GI tract.
V. E XPERIMENTS
In this section, we will provide a detailed evaluation of the
proposed objective function and the robotic platform.
A. Experimental Protocols
1) Datasets: The C3VD 

## B0035 · 第8页 · 清洗文本字符1201–2400

[16] and SimCol dataset [17] were
used for training and evaluation. The upper-GI dataset [19],
EndoSLAM dataset [3], and EndoMapper dataset [18] were
utilised for generalisability evaluation, while the realistic upper
GI tract phantom served as a validation platform for automatic
intervention.
(i) The C3VD dataset [16] is a 3D video colonoscopy
dataset acquired with a high-definition clinical colono-
scope and high-fidelity colon phantoms. It primarily
comprises 20 videos, each of which records segments
of the lower GI tract, including categories such as the
cecum, descending colon, sigmoid colon and transcend-
ing colon.
(ii) The SimCol dataset [17] are sourced from virtual envi-
ronments built with Unity Engine. SimCol contains 33
videos, each recording the entire lower GI tract.
(iii) The upper-GI dataset [19] consists of 3704 images
from 11 landmarks, such as pharynx, middle-upper body,
antrum (see Fig. 9).
(iv) The EndoSLAM dataset [3] encompasses a wide range
of synthetic and real endoscopic video data, capturing the
diverse and challenging environments encountered within
various parts of colons.
(v) The EndoMapper dataset [18] comprises 59 sequences
totalling over 15 hours o

## B0036 · 第8页 · 清洗文本字符2401–3600

f video, providing benchmark
for the evaluation of localisation and navigation of the
endoscope.
(vi) A realistic upper GI tract phantom made of soft
rubber material (see Fig. 5(b)) was used for qualitative
evaluation of automatic intervention.
The data splits for our experiments, including training,
validation, and test sets, are organised as follows:
(i) The C3VD split was arranged as shown in Table I, where
the textures for testing would not be seen during training.
Due to the slow camera movement, we set the triplet
index to (-20, 0, 20). To further remove static frames,
we downsample triplets to 320 × 256, perform feature
matching [15] for consecutive frames in triplet, and
remove those triplets whose average distance between
feature points is less than 1 pixel. We finally got our
C3VD split which has a total of 4,524 monocular triplets
for training, 1,068 for validation and 1,261 for testing.
(ii) The SimCol split was organised as depicted in Table
II, where videos from various sets may share the same
texture. Since the camera speed is fast, we set the triplet
index to (-3, 0, 3). We finally got a total of 21,510
monocular triplets for training, 7,161 for validation and
8,951

## B0037 · 第8页 · 清洗文本字符3601–4800

 for testing.
2) Implementation Details: The training and inference of our
models are supported by two NVIDIA RTX 3090 GPUs and
four Titan Xp GPUs. We summarise our hyperparameters in
Table III following the setting from Monodepth2 [10], Lite-
mono [12], and MonoViT [11]. Unless otherwise specified, all
models on the SimCol dataset have an input/output resolution
of 352 × 352, and all models on the C3VD dataset have an
input/output resolution of 320 × 256. The objective function
of GT depth-supervised baselines is defined as:
L = (1 − α)∥Da − Dgt∥1 + α
2 (1 − SSIM(Da, Dgt)), (25)
where Dgt denotes the GT depth. Lite-mono with FBIS and
PM trained on C3VD is used to generate depth maps for
automatic robotic endoscopy.
TABLE III
DETAIL HYPERPARAMETERS .
Method Epoch Batch size Optimiser Scheduler Initial LR Last LR
Monodepth2 [10] 25 12 Adam StepLR(20,0.1) 10 −4 10−5
Lite-mono [12] 25 12 AdamW CosineAnnealing(·) 10 −4 10−5
MonoViT [11] 25 10 AdamW ExponetialLR(0.9) 5 ×10−5 -
3) Evaluation Metrics: The performance is evaluated by
seven metrics proposed in prior works [31], including mean
absolute relative error (Abs Rel), squared relative error (Sq
Rel), root mean squared error (RMSE),

## B0038 · 第8页 · 清洗文本字符4801–5976

 root mean squared
log error (RMSE log), and the accuracies under threshold
(δ < 1.25, δ < 1.252, δ < 1.253). Following the evaluation
protocol established in [10], we present results using per-
image median GT depth scaling. Additionally, we perform five
training runs for each model, calculating average and standard
deviation values for every evaluation metric. Subsequently,
qualitative evaluation is conducted on the model with median
performance from these five runs (see Figs. 6 and 7). Unless
otherwise noted, all qualitative results are colour-mapped
based on their absolute depth.
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0039 · 第9页 · 清洗文本字符1–1200

QI et al.: MONOLOT: SELF-SUPERVISED MONOCULAR DEPTH ESTIMATION IN LOW-TEXTURE SCENES FOR AUTOMATIC ROBOTIC ENDOSCOPY 9
TABLE IV
COMPARISON WITH BASELINE METHODS ON THE C3VD DATASET. “D": G ROUND TRUTH DEPTH SUPERVISION . “M": S ELF -SUPERVISED MONO
SUPERVISION . THE BEST RESULTS ARE IN BOLD .
Method Train Depth Error(↓) Depth Accuracy( ↑)
Abs rel Sq rel RMSE RMSE log δ <1.25 δ <1.252 δ <1.253
Monodepth2 [10] (baseline) D 0.119 ±0.001 0.006±0.000 0.039±0.001 0.141±0.001 0.877±0.003 0.976±0.001 0.998±0.000
Lite-mono [12] (baseline) D 0.097 ±0.000 0.004±0.000 0.037±0.000 0.120±0.000 0.901±0.000 0.996±0.000 1.000±0.000
MonoViT [11] (baseline) D 0.078±0.000 0.003±0.000 0.033±0.000 0.096±0.000 0.941±0.001 0.999±0.000 1.000±0.000
Monodepth2 [10] (baseline) M 0.160 ±0.004 0.014±0.000 0.080±0.001 0.207±0.004 0.746±0.017 0.945±0.004 0.988±0.004
Lite-mono [12] (baseline) M 0.102 ±0.009 0.007±0.001 0.062±0.006 0.139±0.013 0.897±0.021 0.984±0.006 0.997±0.002
MonoViT [11] (baseline) M 0.114 ±0.008 0.009±0.001 0.070±0.007 0.156±0.015 0.864±0.029 0.978±0.008 0.996±0.003
Monodepth2 + L∗
bis + Lm (MonoLoT) M 0.096 ±0.001 0.005±0.000 0.043±0.001 0.118±0.001 0.934±0.002 0.997±0.001 1.000±0.000
Lite-mo

## B0040 · 第9页 · 清洗文本字符1201–2400

no + L†
bis + Lm (MonoLoT) M 0.088±0.001 0.004±0.000 0.044±0.001 0.111±0.001 0.944±0.001 0.997±0.001 1.000±0.000
MonoViT + L†
bis + Lm (MonoLoT) M 0.090 ±0.002 0.005±0.000 0.046±0.002 0.114±0.002 0.942±0.003 0.997±0.001 1.000±0.000
TABLE V
COMPARISON WITH BASELINE METHODS ON THE SIMCOL DATASET . “D": G ROUND TRUTH DEPTH SUPERVISION . “M": S ELF -SUPERVISED MONO
SUPERVISION . GROUND TRUTH DEPTH VALUES RESCALED TO [0, 20] CM. THE BEST RESULTS ARE IN BOLD .
Method Train Depth Error(↓) Depth Accuracy( ↑)
Abs rel Sq rel RMSE RMSE log δ <1.25 δ <1.252 δ <1.253
Monodepth2 [10] (baseline) D 0.030 ±0.000 0.005±0.000 0.118±0.001 0.041±0.000 0.996±0.000 0.999±0.000 1.000±0.000
Lite-mono [12] (baseline) D 0.055 ±0.000 0.041±0.000 0.194±0.000 0.064±0.000 0.987±0.000 0.997±0.000 0.999±0.000
MonoViT [11] (baseline) D 0.025±0.000 0.004±0.000 0.108±0.002 0.035±0.000 0.997±0.000 0.999±0.000 1.000±0.000
Monodepth2 [10] (baseline) M 0.160 ±0.022 1.008±0.306 0.724±0.084 0.169±0.016 0.907±0.008 0.953±0.004 0.970±0.004
Lite-mono [12] (baseline) M 0.171 ±0.008 1.318±0.119 0.742±0.026 0.167±0.004 0.913±0.003 0.951±0.001 0.967±0.001
MonoViT [11] (baseline) M 0.143 ±0.014 0.562±0.182 0.627±0.050 0.171±0.008 

## B0041 · 第9页 · 清洗文本字符2401–3600

0.901±0.006 0.951±0.003 0.969±0.002
Monodepth2 + L∗
bis + Lm (MonoLoT) M 0.074 ±0.002 0.066±0.000 0.370±0.006 0.104±0.002 0.959±0.002 0.988±0.000 0.995±0.001
Lite-mono + L∗
bis + Lm (MonoLoT) M 0.075 ±0.000 0.069±0.000 0.394±0.003 0.107±0.001 0.957±0.001 0.987±0.000 0.994±0.000
MonoViT + L†
bis + Lm (MonoLoT) M 0.073±0.002 0.065±0.001 0.394±0.008 0.105±0.002 0.958±0.001 0.988±0.000 0.995±0.000
TABLE VI
ABLATION STUDIES OF FBIS, RBIS AND PM ON THE C3VD DATASET. THE BEST RESULTS ARE IN BOLD . THE SECOND BEST ARE UNDERLINED .
DepthNet Lbis† Lbis* Lm
Depth Error(↓) Depth Accuracy( ↑)
Abs rel Sq rel RMSE RMSE log δ <1.25 δ <1.252 δ <1.253
Monodepth2 [10]
(baseline) 0.160 ±0.004 0.014±0.000 0.080±0.001 0.207±0.004 0.746±0.017 0.945±0.004 0.988±0.004
✓ 0.103±0.002 0.006±0.000 0.053±0.001 0.132±0.003 0.905±0.008 0.993±0.002 1.000±0.000
✓ 0.098±0.002 0.005±0.000 0.046±0.001 0.121±0.002 0.928±0.005 0.996±0.000 1.000±0.000
✓ 0.147±0.009 0.012±0.002 0.069±0.004 0.187±0.011 0.791±0.019 0.947±0.010 0.989±0.006
✓ ✓ 0.098±0.003 0.006±0.000 0.051±0.001 0.124±0.002 0.920±0.004 0.995±0.001 1.000±0.000
✓ ✓ 0.096±0.001 0.005±0.000 0.043±0.001 0.118±0.001 0.934±0.002 0.997±0.001 1.000±0.000
Lite-mono [1

## B0042 · 第9页 · 清洗文本字符3601–4800

2]
(baseline) 0.102 ±0.009 0.007±0.001 0.062±0.006 0.139±0.013 0.897±0.021 0.984±0.006 0.997±0.002
✓ 0.093±0.003 0.005±0.000 0.048±0.001 0.118±0.002 0.935±0.004 0.996±0.000 1.000±0.000
✓ 0.102±0.002 0.006±0.000 0.051±0.001 0.126±0.002 0.920±0.005 0.996±0.000 1.000±0.000
✓ 0.086±0.002 0.005±0.000 0.046±0.001 0.112±0.002 0.933±0.007 0.995±0.002 1.000±0.000
✓ ✓ 0.088±0.001 0.004±0.000 0.044±0.001 0.111±0.001 0.944±0.001 0.997±0.001 1.000±0.000
✓ ✓ 0.092±0.001 0.005±0.000 0.046±0.001 0.116±0.001 0.938±0.002 0.997±0.000 1.000±0.000
MonoViT [11]
(baseline) 0.114 ±0.008 0.009±0.001 0.070±0.007 0.156±0.015 0.864±0.029 0.978±0.008 0.996±0.003
✓ 0.116±0.004 0.008±0.001 0.062±0.002 0.148±0.005 0.872±0.011 0.990±0.001 0.999±0.001
✓ 0.134±0.004 0.011±0.001 0.071±0.002 0.169±0.004 0.819±0.011 0.986±0.002 0.998±0.001
✓ 0.092±0.007 0.006±0.001 0.057±0.005 0.126±0.010 0.919±0.014 0.984±0.003 0.996±0.002
✓ ✓ 0.090±0.001 0.005±0.000 0.046±0.002 0.114±0.002 0.943±0.003 0.997±0.001 1.000±0.000
✓ ✓ 0.120±0.010 0.009±0.002 0.064±0.008 0.153±0.014 0.862±0.031 0.988±0.004 0.998±0.001
B. Quantitative Results
Results comparing with the self-supervised baseline and
the supervised baseline are reported in Tabl

## B0043 · 第9页 · 清洗文本字符4801–6000

es IV and V. To
further validate the effectiveness of MonoLoT, an ablation
study was conducted to assess the significance of various
design components within the architecture (refer to Table VI
and Table VII). The results are reported for the C3VD and
SimCol datasets.
C. Qualitative Results
A comprehensive qualitative evaluation of MonoLoT is
organised as follows:
(i) Results for the high-definition phantom ( C3VD) [16] and
virtual colonoscopy (SimCol) [17] are presented in Figs.
6 and 7, respectively.
(ii) The generalisation capability of MonoLoT on real GI
endoscopy is evaluated using the upper-GI dataset [19]
(see Fig 9), the EndoSLAM dataset, and the EndoMapper
dataset (see Fig 8).
(iii) Steering capabilities consists linear trajectories along
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.


## B0044 · 第9页 · 清洗文本字符6001–6143

Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0045 · 第10页 · 清洗文本字符1–1200

10 IEEE TRANSACTIONS AND JOURNALS TEMPLATE
TABLE VII
ABLATION STUDIES OF FBIS, RBIS AND PM ON THE SIMCOL DATASET . THE BEST RESULTS ARE IN BOLD . THE SECOND BEST ARE UNDERLINED .
DepthNet Lbis† Lbis* Lm
Depth Error(↓) Depth Accuracy( ↑)
Abs rel Sq rel RMSE RMSE log δ <1.25 δ <1.252 δ <1.253
Monodepth2 [10]
(baseline) 0.160 ±0.022 1.008±0.306 0.724±0.084 0.169±0.016 0.907±0.008 0.953±0.004 0.970±0.004
✓ 0.074±0.000 0.060±0.001 0.365±0.001 0.103±0.000 0.962±0.001 0.989±0.000 0.995±0.000
✓ 0.078±0.001 0.068±0.001 0.370±0.005 0.106±0.001 0.957±0.001 0.987±0.000 0.994±0.000
✓ 0.124±0.001 0.613±0.013 0.522±0.005 0.140±0.001 0.937±0.000 0.976±0.001 0.987±0.000
✓ ✓ 0.084±0.002 0.098±0.020 0.403±0.009 0.115±0.002 0.952±0.001 0.985±0.001 0.992±0.001
✓ ✓ 0.074±0.002 0.066±0.000 0.370±0.006 0.104±0.002 0.959±0.002 0.988±0.000 0.995±0.001
Lite-mono [12]
(baseline) 0.171 ±0.008 1.318±0.119 0.742±0.026 0.167±0.004 0.913±0.003 0.951±0.001 0.967±0.001
✓ 0.080±0.001 0.073±0.001 0.400±0.003 0.112±0.001 0.954±0.001 0.985±0.000 0.993±0.000
✓ 0.084±0.003 0.079±0.004 0.406±0.006 0.115±0.003 0.950±0.003 0.984±0.000 0.993±0.000
✓ 0.117±0.002 0.640±0.045 0.491±0.003 0.133±0.001 0.943±0.001 0.979±0.000 0.988

## B0046 · 第10页 · 清洗文本字符1201–2400

±0.000
✓ ✓ 0.078±0.000 0.069±0.001 0.401±0.003 0.110±0.000 0.954±0.001 0.986±0.000 0.994±0.000
✓ ✓ 0.075±0.000 0.069±0.000 0.394±0.003 0.107±0.001 0.957±0.001 0.987±0.000 0.994±0.000
MonoViT [11]
(baseline) 0.143 ±0.014 0.562±0.182 0.627±0.050 0.171±0.008 0.901±0.006 0.951±0.003 0.969±0.002
✓ 0.085±0.002 0.074±0.001 0.402±0.006 0.113±0.002 0.952±0.002 0.986±0.000 0.994±0.000
✓ 0.093±0.002 0.081±0.002 0.424±0.012 0.121±0.002 0.943±0.003 0.984±0.001 0.993±0.000
✓ 0.128±0.002 0.796±0.069 0.533±0.011 0.139±0.001 0.941±0.001 0.977±0.001 0.986±0.000
✓ ✓ 0.073±0.002 0.065±0.001 0.394±0.008 0.105±0.002 0.958±0.001 0.988±0.000 0.995±0.000
✓ ✓ 0.076±0.001 0.070±0.000 0.401±0.005 0.108±0.001 0.956±0.001 0.987±0.000 0.994±0.000
the X axis, Y axis, and two diagonal lines with angles
ranging from ±40°to ±240°, and circular trajectories
with radii ranging from 40°to 160°, all in 40° increments
(see Fig 10).
(iv) The automatic robotic endoscopy on the realistic
upper GI tract phantom showcases the user interface,
GT pose in 3D space, and the appearance of the robotic
platform during insertion and retraction (see Video S1).
The insertion or retraction process encompasses five
stages (see Fig. 1).
V

## B0047 · 第10页 · 清洗文本字符2401–3600

I. D ISCUSSION
This study aimed to improve the performance of self-
supervised monocular depth estimation methods in real-world
endoscopic scenarios with low-texture regions. Furthermore,
we designed and implemented a robotic platform for control-
ling digestive endoscopy and ultimately validated the feasibil-
ity of using the proposed monocular depth estimation method
for automatic robotic endoscopy. In this section, we will
provide a detailed analysis of the proposed objective functions
and the robotic platform.
A. Performance of the Monocular Depth Estimation
Typically, there exists a performance gap between self-
supervised methods and GT depth-supervised methods [10],
as well as between CNN-based methods and Transformer-
based methods [11], [32]. However, our experiments on low-
texture scenes suggest an exception, as they demonstrate that
self-supervised methods can achieve competitive performance
comparable to GT depth-supervised methods. Moreover, the
disparity between CNN-based methods and Transformer-based
methods can be mitigated.
1) C3VD Dataset: As demonstrated in Table IV, the “base-
lines (M)” exhibited notably lower accuracy compared to
the “baselines (D)”, particul

## B0048 · 第10页 · 清洗文本字符3601–4800

arly when evaluating the “ δ <
1.25” metric. However, the performance of MonoLoT-enabled
models, namely “MonoLoT (M)”, improved significantly, even
surpassing the “baselines (D)”. For instance, the “ δ < 1.25”
metric of “Monodepth2 (M)” increased from 0.746 (baseline)
to 0.934 (MonoLoT), surpassing that of “Monodepth2 (D)”
(0.877). A similar trend is observable for Lite-mono and
MonoViT. In this case, the performance gap of the “ δ <1.25”
between “Monodepth2 (M)” and “Lite-mono (M)” reduced
from 0.118 (baseline) to 0.01 (MonoLoT).
Comprehensive ablation studies are reported in Table VI,
highlighting that a majority of the combinations involving
the proposed batch image shuffle and point matching loss
yield significant performance improvements for the baselines.
Notably, the combination of RBIS and PM demonstrated
superior performance on Monodepth2, while the combination
of FBIS and PM outperformed other combinations on Lite-
mono and MonoViT.
Additional qualitative results for the C3VD dataset (see
Fig. 6) reveal a notable side effect of FBIS on Monodepth2.
This effect is characterised by the presence of a borderline
running from top to bottom and another border line from
left to r

## B0049 · 第10页 · 清洗文本字符4801–6000

ight in the depth estimation. Fortunately, this issue
was effectively mitigated by employing RBIS. Such an effect
was not observed in Transformer-based methods, such as Lite-
mono and MonoViT. Experimental results showed that, in
most case, FBIS brought more benefit to Transformer-based
methods compared to RBIS.
2) SimCol Dataset: A similar trend of improved perfor-
mance is observed in Table V on the SimCol dataset. By
using the proposed loss function, the “RMSE” metric of
all the ”baseline (M)” models was reduced by half, while
the “ δ < 1.25” metric increased by around 5%. We also
conduct extensive experiments of ablation studies on the
SimCol dataset as reported in Table VII, which shows all the
combinations of proposed loss functions brought increasing
performance. The combination with the best performance are
selected as our final models for comparison with the baseline
in Table V, except the “Monodepth2 + L†
bis” model because of
the side effect discussed before. In alternative, the second best
results of Monodepth2 in Table VII, namely “Monodepth2 +
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's vers

## B0050 · 第10页 · 清洗文本字符6001–6459

ion which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0051 · 第11页 · 清洗文本字符1–1200

QI et al.: MONOLOT: SELF-SUPERVISED MONOCULAR DEPTH ESTIMATION IN LOW-TEXTURE SCENES FOR AUTOMATIC ROBOTIC ENDOSCOPY 11
ℒ𝑚 ℒ𝑏𝑖𝑠
† + ℒ𝑚 ℒ𝑏𝑖𝑠
∗ + ℒ𝑚ℒ𝑏𝑖𝑠
∗ℒ𝑏𝑖𝑠
†baseline
Monodepth2Lite-monoMonoViT
RGB
Ground truth
Cecum
0 1
Monodepth2Lite-monoMonoViT
Descending
Ground truth
RGB
0 1
Monodepth2Lite-monoMonoViT
Sigmoid
Ground truth
RGB
0 1
Monodepth2Lite-monoMonoViT
Transcending
Ground truth
RGB
0 1
Fig. 6. Qualitative C3VD [16] results. Monodepth2 + L∗
bis + Lm, Lite-
mono + L†
bis + Lm, and MonoViT + L†
bis + Lm exhibit equally excel-
lent results, outperforming the others. The RMSE values are presented
below each depth map, with the best results highlighted in bold.
L∗
bis + Lm”, showed competing performance to “Monodepth2
+ L†
bis” was taken as the best model.
The qualitative results depicted in Fig. 7 provide instances
where the baseline models encountered failures. However,
“SynCol I” in Fig. 7 represents a distinct failure case unrelated
to the baseline models, attributable to the lighting setup.
Specifically, the lower-left part of the image, where the diges-
tive tract obstructs one of the light sources, results in a dark
region. It’s noteworthy that the SimCol dataset employs t

## B0052 · 第11页 · 清洗文本字符1201–2400

wo
light sources, one on each side of the camera. Consequently,
when one light source is obstructed, the other remains work.
In contrast, real endoscopes typically feature a single light
source. If this light source becomes obstructed, it leads to two
potential outcomes: firstly, a strong reflection may emerge,
causing the obstructed area to appear bright. Secondly, most
of the light rays may be entirely blocked, resulting in reduced
overall brightness in the image. Consequently, we posit that the
lighting setup in the SimCol dataset could potentially impact
0 20
Monodepth2Lite-monoMonoViT
RGB
Ground truth
SynCol I
SynCol II
SynCol III
ℒ𝑚 ℒ𝑏𝑖𝑠
† + ℒ𝑚 ℒ𝑏𝑖𝑠
∗ + ℒ𝑚ℒ𝑏𝑖𝑠
∗ℒ𝑏𝑖𝑠
†baseline
0 20
RGB
Ground truth
0 20
RGB
Ground truth
Monodepth2Lite-monoMonoViTMonodepth2Lite-monoMonoViT
Fig. 7. Qualitative SimCol [17] results. GT depth values rescaled to
[0, 20] cm. Monodepth2 + L∗
bis + Lm, Lite-mono + L∗
bis + Lm, and
MonoViT + L†
bis + Lm exhibit equally excellent results, outperforming
the others. The RMSE values are presented below each depth map, with
the best results highlighted in bold.
the model’s generalisation to real endoscopic images, as the
lighting condition is a critical fact

## B0053 · 第11页 · 清洗文本字符2401–3600

or in determining depth.
3) Generalisability of the Model: In general, models trained
on the C3VD dataset exhibited greater robustness to reflected
light compared to those trained on the SimCol dataset, as
the illumination setting and texture of the C3VD dataset
closely resemble real endoscopy conditions (see Fig. 9). The
predicted depth maps of Monodepth2 are similar to those of
Lite-mono and MonoViT when models are trained with the
proposed objective function, thus validating that MonoLoT
has reduced the performance gap between CNN-based meth-
ods and Transformer-based methods. Moreover, MonoLoT-
enabled methods trained on the C3VD dataset produced
more accurate range of depth than the others. For example,
MonoLoT-enabled self-supervised Monodepth2, Lite-mono,
and MonoViT trained on C3VD produced accurate depth maps
showed that the middle-upper body in Fig. 9 covers a wider
range of depth than the others. “Lite-mono + L†
bis + Lm”
model trained on C3VD demonstrated strong generalisability
and was subsequently integrated to the robotic platform. Ad-
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version whi

## B0054 · 第11页 · 清洗文本字符3601–4052

ch has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0055 · 第12页 · 清洗文本字符1–1200

12 IEEE TRANSACTIONS AND JOURNALS TEMPLATE
NA NA NA
RGBGround TruthMonoLoT
NA NANA
EndoSLAM EndoMapper
Fig. 8. Zero-shot evaluation on EndoSLAM and EndoMapper datasets demonstrates MonoLoT’s capability to generate depth maps that accurately
capture spatial structures and exhibit robustness to reflections for unseen synthetic and real images.
0 1
Monodepth2Lite-monoMonoViT
RGB
Pharynx
0 1
Monodepth2Lite-monoMonoViT
RGB
Middle-
upper
body
0 1
Monodepth2Lite-monoMonoViT
RGB
Antrum
Trained on C3VD Trained on SimCol
Depth
supervised Self-supervised Depth
supervised Self-supervised
baseline baseline MonoLoT baseline baseline MonoLoT
Fig. 9. Qualitative results showing the prediction of selected anatomical
landmarks in upper-GI images [19]. MonoLoT -enabled methods trained
on C3VD showed an accurate range of depth and robustness to re-
flected light. Predicted depths of models trained on the SimCol dataset
are clipped to the range [0, 10] cm and then resized to the range [0, 1]
for alignment with the C3VD dataset.
ditionally, we evaluated MonoLoT on the EndoSLAM dataset
and the EndoMapper dataset under a zero-shot setting, as
shown in Fig. 8. Since the intrinsic matrix of the training set

## B0056 · 第12页 · 清洗文本字符1201–2400


and test set are different, the predicted depths are inevitably
different from the ground truth. Due to discrepancies in
the intrinsic matrices between the training and test sets, the
predicted depths diverge from the ground truth. However, the
qualitative outcomes indicate that our approach successfully
captures accurate details across a wide range of endoscopic
videos.
(c)
X [mm]
Y [mm]
(d)
X [mm]
Y [mm]
(e)
Angle of the 2nd motor [deg]
Angle of the 1st motor [deg]
(f)
Angle of the 2nd motor [deg]
Angle of the 1st motor [deg]
Z [mm]
(a)
Z [mm]
(b)
Fig. 10. Visualisation of linear trajectories (a, c) and circular trajectories
(b, d) in 3D space (a, b) and the X-Y plane (c, d), accompanied by the
corresponding motor degrees (e, f). The initiation and termination points
of each linear trajectory are marked with timestamps and ⋆ symbols.
B. Performance of the Robotic Platform
1) Mechanical Structure: The proposed robotic platform
provides a concise interface to the Olympus endoscopy system
and is portable enough to be adapted for use with similar
endoscopy systems due to its simplified design (see Fig. 5
(a)). Considering the potential motion errors introduced by the
soft endoscope 

## B0057 · 第12页 · 清洗文本字符2401–3220

mechanism and forces between the endoscope
and tissues, it is essential to assess their impact on endoscopic
kinematics. Fortunately, we have found that PID control is a
viable alternative. Experimental results for linear and circular
This article has been accepted for publication in IEEE Journal of Biomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0058 · 第13页 · 清洗文本字符1–1200

QI et al.: MONOLOT: SELF-SUPERVISED MONOCULAR DEPTH ESTIMATION IN LOW-TEXTURE SCENES FOR AUTOMATIC ROBOTIC ENDOSCOPY 13
motions using PID control in an open space have shown that
the trajectories of the endoscopic tip closely follow the motion
patterns of the motors (see Fig. 10). Though relocation error
is unavoidable, the depth-based PID control can effectively
address this issue.
2) Depth-based PID Control for Automatic Intervention: To
illustrate the real-time applicability of depth-based PID control
in our robotic platform, we present a test video showcasing
the insertion and retraction procedures in a realistic upper GI
phantom. Full details can be seen in the supplementary materi-
als (Video S1). MonoLoT has demonstrated excellent general-
isability on a realistic upper GI phantom (see Fig. 1), providing
effective support for the anchor point computation. Unlike [7]
and [8], our controller prototype is designed to be adaptable to
existing endoscopes and robotic platforms, regardless of their
structure, as long as they offer appropriate motion interfaces.
We utilised the PID controller for the movement of the
bending section because this model-free controller is robust
and wi

## B0059 · 第13页 · 清洗文本字符1201–2400

dely used in the same application, such as [33]. The PID
controller is more suitable in this scenario than the Jacobian
controller because the commercial endoscope already provides
a control interface, and it is not possible to directly drive the
sheath to control the endoscope kinematically. Therefore, we
ultimately use the PID controller to control the movement
of the endoscope. Guided by an anchor point determined
through depth estimation, the PID controller systematically
navigates the endoscope from the pharynx to the antrum.
Our experimental findings robustly demonstrate the feasibility
and effectiveness of depth-based PID control for autonomous
interventions in digestive endoscopy. Consequently, this pro-
posed method holds significant promise for clinical application
in automated gastrointestinal examinations, thereby liberating
endoscopists to concentrate more on critical decision-making
processes. Furthermore, intricate endoscopic procedures such
as endoscopic retrograde cholangiopancreatography (ERCP)
and endoscopic submucosal dissection (ESD), which typically
require multiple operators for manual endoscopic interventions
and handling of surgical instruments, could benef

## B0060 · 第13页 · 清洗文本字符2401–3600

it from the
introduction of an automated robotic system. This system has
the potential to mitigate issues related to miscommunication,
the allocation of costly medical resources, and extended dura-
tions of ERCP and ESD procedures by assisting surgeons in
the execution of these endoscopic tasks.
C. Limitations and Future Works
Our experimental results expose some shortcomings of
our robot design. As the intervention distance increases, the
deformation of endoscope also increases due to the increasing
contact between the endoscope and the gastrointestinal tract,
leading to difficulty in inserting the endoscope further. This
can be solved with custom-made endoscopes with variable
stiffness and additional manipulators [34]. Furthermore, au-
tonomous endoscopic intervention requires safe interaction
with the GI tract, thus an additional interactive force mea-
surement helps ensure the intervention’s safety [35]. On the
other hand, our vision-based algorithm overcomes the low-
texture challenge, but still shares some limitations similar to
the other self-supervised depth estimation algorithm, incapable
of predicting the absolute depth of images in various scenes.
Fortunately, the latest

## B0061 · 第13页 · 清洗文本字符3601–4800

 trends based on few-shot learning and
pre-trained vision language model [36] shows a potential to
address this issue. Moreover, the predicted depth around the
vanishing point of the GI image shows a counter-intuitive
closed surface that deserves to be explored in future work.
VII. C ONCLUSION
Our research has introduced an innovative approach that ad-
dresses the challenges associated with self-supervised monoc-
ular depth estimation in digestive endoscopy. We have ad-
dressed two critical aspects: the limitations of self-supervised
depth estimation in low-texture scenes and the application of
depth estimation in visual servoing for digestive endoscopy.
Our investigation has revealed that the struggles of self-
supervised depth estimation in low-texture scenes stem from
inaccurate photometric reconstruction losses. To overcome
this, we have introduced the point-matching loss, which refines
the reprojected points. Furthermore, during the training pro-
cess, data augmentation is achieved through batch image shuf-
fle loss, significantly improving the accuracy and generalisa-
tion capability of the depth model. The combined contributions
of the point matching loss and batch image shu

## B0062 · 第13页 · 清洗文本字符4801–6000

ffle loss have
boosted the baseline accuracy by a minimum of 5% on both
the C3VD and SimCol datasets, surpassing the generalisability
of ground truth depth-supervised baselines when applied to
upper-GI datasets. Moreover, the successful implementation
of our robotic platform for automatic intervention in digestive
endoscopy demonstrates the practical and impactful applica-
tion of monocular depth estimation technology.
In conclusion, our work not only advances the field of en-
doscopic image analysis but also holds promise for enhancing
clinical procedures and ultimately improving patient outcomes
in gastrointestinal healthcare.
REFERENCES
[1] Q. He, S. Bano, O. F. Ahmad, B. Yang, X. Chen, P. Valdastri, L. B.
Lovat, D. Stoyanov, and S. Zuo, “Deep learning-based anatomical
site classification for upper gastrointestinal endoscopy,” International
Journal of Computer Assisted Radiology and Surgery , vol. 15, no. 7,
pp. 1085–1094, Jul. 2020.
[2] A. R. Widya, Y . Monno, M. Okutomi, S. Suzuki, T. Gotoda, and K. Miki,
“Whole stomach 3d reconstruction and frame localization from monoc-
ular endoscope video,” IEEE Journal of Translational Engineering in
Health and Medicine , vol. 7, pp. 1–10

## B0063 · 第13页 · 清洗文本字符6001–7200

, 2019.
[3] K. B. Ozyoruk, G. I. Gokceler, T. L. Bobrow, G. Coskun, K. Incetan,
Y . Almalioglu, F. Mahmood, E. Curto, L. Perdigoto, M. Oliveira et al.,
“Endoslam dataset and an unsupervised monocular visual odometry
and depth estimation approach for endoscopic videos,” Medical image
analysis, vol. 71, p. 102058, 2021.
[4] G. Ciuti, K. Skonieczna- ˙Zydecka, W. Marlicz, V . Iacovacci, H. Liu,
D. Stoyanov, A. Arezzo, M. Chiurazzi, E. Toth, H. Thorlacius et al. ,
“Frontiers of robotic colonoscopy: a comprehensive review of robotic
colonoscopes and technologies,” Journal of clinical medicine , vol. 9,
no. 6, p. 1648, 2020.
[5] F. Chadebecq, L. B. Lovat, and D. Stoyanov, “Artificial intelligence and
automation in endoscopy and surgery,”Nature Reviews Gastroenterology
& Hepatology, vol. 20, no. 3, pp. 171–182, 2023.
[6] A. Pore, M. Finocchiaro, D. Dall’Alba, A. Hernansanz, G. Ciuti,
A. Arezzo, A. Menciassi, A. Casals, and P. Fiorini, “Colonoscopy
navigation using end-to-end deep visuomotor control: A user study,”
in 2022 IEEE/RSJ International Conference on Intelligent Robots and
Systems (IROS). IEEE, 2022, pp. 9582–9588.
This article has been accepted for publication in IEEE Journal of B

## B0064 · 第13页 · 清洗文本字符7201–7718

iomedical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.
Authorized licensed use limited to: University College London. Downloaded on July 18,2024 at 15:31:36 UTC from IEEE Xplore. Restrictions apply.

## B0065 · 第14页 · 清洗文本字符1–1200

14 IEEE TRANSACTIONS AND JOURNALS TEMPLATE
[7] T. Yang, Y . Yang, P. Wang, Y . Cao, Z. Yang, and H. Liu, “A lumen-
adapted navigation scheme with spatial awareness from monocular
vision for autonomous robotic endoscopy,” Robotics and Autonomous
Systems, vol. 165, p. 104444, 2023.
[8] L. Ott, F. Nageotte, P. Zanne, and M. De Mathelin, “Robotic assistance
to flexible endoscopy by physiological-motion tracking,” IEEE Trans-
actions on Robotics , vol. 27, no. 2, pp. 346–359, 2011.
[9] L. Barducci, B. Scaglioni, J. Martin, K. L. Obstein, and P. Valdastri,
“Active stabilization of interventional tasks utilizing a magnetically
manipulated endoscope,” Frontiers in Robotics and AI, vol. 9, p. 854081,
2022.
[10] C. Godard, O. Mac Aodha, M. Firman, and G. J. Brostow, “Digging
into self-supervised monocular depth estimation,” in Proceedings of the
IEEE/CVF international conference on computer vision, 2019, pp. 3828–
3838.
[11] C. Zhao, Y . Zhang, M. Poggi, F. Tosi, X. Guo, Z. Zhu, G. Huang,
Y . Tang, and S. Mattoccia, “Monovit: Self-supervised monocular depth
estimation with a vision transformer,” in 2022 International Conference
on 3D Vision (3DV) , 2022, pp. 668–678.
[12] N. Zhang, F. Nex, G

## B0066 · 第14页 · 清洗文本字符1201–2400

. V osselman, and N. Kerle, “Lite-mono: A
lightweight cnn and transformer architecture for self-supervised monoc-
ular depth estimation,” in Proceedings of the IEEE/CVF Conference on
Computer Vision and Pattern Recognition , 2023, pp. 18 537–18 546.
[13] A. Rau, P. E. Edwards, O. F. Ahmad, P. Riordan, M. Janatka, L. B.
Lovat, and D. Stoyanov, “Implicit domain adaptation with conditional
generative adversarial networks for depth prediction in endoscopy,” In-
ternational journal of computer assisted radiology and surgery , vol. 14,
pp. 1167–1176, 2019.
[14] J. Watson, O. Mac Aodha, V . Prisacariu, G. Brostow, and M. Fir-
man, “The temporal opportunist: Self-supervised multi-frame monocular
depth,” in Proceedings of the IEEE/CVF Conference on Computer Vision
and Pattern Recognition, 2021, pp. 1164–1174.
[15] J. Sun, Z. Shen, Y . Wang, H. Bao, and X. Zhou, “Loftr: Detector-
free local feature matching with transformers,” in Proceedings of the
IEEE/CVF conference on computer vision and pattern recognition, 2021,
pp. 8922–8931.
[16] T. L. Bobrow, M. Golhar, R. Vijayan, V . S. Akshintala, J. R. Garcia,
and N. J. Durr, “Colonoscopy 3d video dataset with paired depth from
2d-3d registration

## B0067 · 第14页 · 清洗文本字符2401–3600

,” arXiv preprint arXiv:2206.08903 , 2022.
[17] A. Rau, B. Bhattarai, L. Agapito, and D. Stoyanov, “Bimodal camera
pose prediction for endoscopy,” IEEE Transactions on Medical Robotics
and Bionics, vol. 5, no. 4, pp. 978–989, 2023.
[18] P. Azagra, C. Sostres, ´A. Ferr ´andez, L. Riazuelo, C. Tomasini, O. L.
Barbed, J. Morlana, D. Recasens, V . M. Batlle, J. J. G ´omez-Rodr´ıguez
et al. , “Endomapper dataset of complete calibrated endoscopy proce-
dures,” Scientific Data, vol. 10, no. 1, p. 671, 2023.
[19] Q. He, S. Bano, O. F. Ahmad, B. Yang, X. Chen, P. Valdastri, L. B.
Lovat, D. Stoyanov, and S. Zuo, “Deep learning-based anatomical site
classification for upper gastrointestinal endoscopy,”International journal
of computer assisted radiology and surgery , vol. 15, pp. 1085–1094,
2020.
[20] A. R. Widya, Y . Monno, M. Okutomi, S. Suzuki, T. Gotoda, and K. Miki,
“Learning-based depth and pose estimation for monocular endoscope
with loss generalization,” in 2021 43rd Annual International Conference
of the IEEE Engineering in Medicine & Biology Society (EMBC). IEEE,
2021, pp. 3547–3552.
[21] J. Rodriguez-Puigvert, D. Recasens, J. Civera, and R. Martinez-Cantin,
“On the uncertain single

## B0068 · 第14页 · 清洗文本字符3601–4800

-view depths in colonoscopies,” in Medical
Image Computing and Computer Assisted Intervention–MICCAI 2022:
25th International Conference, Singapore, September 18–22, 2022,
Proceedings, Part III. Springer, 2022, pp. 130–140.
[22] B. Li, Y . Huang, Z. Liu, D. Zou, and W. Yu, “Structdepth: Leveraging
the structural regularities for self-supervised indoor depth estimation,”
in Proceedings of the IEEE/CVF International Conference on Computer
Vision, 2021, pp. 12 663–12 673.
[23] K. Saunders, G. V ogiatzis, and L. J. Manso, “Self-supervised monocular
depth estimation: Let’s talk about the weather,” in Proceedings of the
IEEE/CVF International Conference on Computer Vision , 2023, pp.
8907–8917.
[24] C. Zhao, M. Poggi, F. Tosi, L. Zhou, Q. Sun, Y . Tang, and S. Mattoccia,
“Gasmono: Geometry-aided self-supervised monocular depth estimation
for indoor scenes,” in Proceedings of the IEEE/CVF International
Conference on Computer Vision , 2023, pp. 16 209–16 220.
[25] Z. Liu, R. Li, S. Shao, X. Wu, and W. Chen, “Self-supervised monocular
depth estimation with self-reference distillation and disparity offset
refinement,” IEEE Transactions on Circuits and Systems for Video
Technology, vol. 33, n

## B0069 · 第14页 · 清洗文本字符4801–6000

o. 12, pp. 7565–7577, 2023.
[26] A. Vaswani, N. Shazeer, N. Parmar, J. Uszkoreit, L. Jones, A. N. Gomez,
Ł. Kaiser, and I. Polosukhin, “Attention is all you need,” Advances in
neural information processing systems , vol. 30, 2017.
[27] Y . Lee, J. Kim, J. Willette, and S. J. Hwang, “Mpvit: Multi-path vision
transformer for dense prediction,” in Proceedings of the IEEE/CVF
Conference on Computer Vision and Pattern Recognition , 2022, pp.
7287–7296.
[28] L.-C. Chen, G. Papandreou, F. Schroff, and H. Adam, “Rethinking
atrous convolution for semantic image segmentation,” arXiv preprint
arXiv:1706.05587, 2017.
[29] A. Ali, H. Touvron, M. Caron, P. Bojanowski, M. Douze, A. Joulin,
I. Laptev, N. Neverova, G. Synnaeve, J. Verbeek et al. , “Xcit: Cross-
covariance image transformers,” Advances in neural information pro-
cessing systems, vol. 34, pp. 20 014–20 027, 2021.
[30] Z. Wang, “Image quality assessment: from error visibility to structural
similarity,” IEEE transactions on image processing , vol. 13, no. 4, pp.
600–612, 2004.
[31] D. Eigen, C. Puhrsch, and R. Fergus, “Depth Map Prediction from a
Single Image using a Multi-Scale Deep Network,” in Advances in Neural
Information Processi

## B0070 · 第14页 · 清洗文本字符6001–7200

ng Systems, Z. Ghahramani, M. Welling, C. Cortes,
N. Lawrence, and K. Q. Weinberger, Eds., vol. 27. Curran Associates,
Inc., 2014.
[32] Y . Liu and S. Zuo, “Self-supervised monocular depth estimation
for gastrointestinal endoscopy,” Computer Methods and Programs in
Biomedicine, p. 107619, 2023.
[33] S. Zuo, T. Chen, X. Chen, and B. Chen, “A wearable hands-free human-
robot interface for robotized flexible endoscope,” IEEE Robotics and
Automation Letters, vol. 7, no. 2, pp. 3953–3960, 2022.
[34] T. Liang, C. Zhang, Y . Wang, K. Kong, X. Chen, B. Wei, S. Wang,
and S. Zuo, “A novel miniature flexible robotic system for endoscopic
mucosal dissection: an animal experimental study,” Journal of Robotic
Surgery, vol. 18, no. 1, pp. 1–7, 2024.
[35] Z. Zhang, C. Zhang, and S. Zuo, “A novel bioinspired whisker sensor for
gastrointestinal endoscopy,” IEEE/ASME Transactions on Mechatronics,
2023.
[36] X. Hu, C. Zhang, Y . Zhang, B. Hai, K. Yu, and Z. He, “Learning to
adapt clip for few-shot monocular depth estimation,” in Proceedings of
the IEEE/CVF Winter Conference on Applications of Computer Vision ,
2024, pp. 5594–5603.
This article has been accepted for publication in IEEE Journal of Biome

## B0071 · 第14页 · 清洗文本字符7201–7714

dical and Health Informatics. This is the author's version which has not been fully edited and
content may change prior to final publication. Citation information: DOI 10.1109/JBHI.2024.3423791
© 2024 IEEE. Personal use is permitted, but republication/redistribution requires IEEE permission.
See https://www.ieee.org/publications/rights/index.html for more information.